Software

- MODbus master for max. 14 units - Memory for 2 mio. measurements - Build-in clock/calender - 10 years battery backup on clock/calender - Data logging from max. 60 nodes via MODbus - 4 analogue inputs, 0(4)-20mA or 0-10V - Galvanic isolation between input and output - Serial transfer of data to PC via RS232

- Measuring range: 0-14pH, ±1000mV, 0-150oC - Single- or dual slope calibration of pH-electrode - Temperature compensation via Pt100/Pt1000 sensor - Readout of pH/mV, temperature - Analogue output 0(4) - 20mA for pH/mV or oC - Invertable analogue output for feedback loop - Galvanic isolation between input and output - Serial communication via MODbus

- Measuring range: 120nA, 1200nA Other ranges on request - Temperature compensation via NTC, 22kohm - Readout of D.O mg/l, % - Analogue output 0(4) - 20mA for oxygen or oC - Invertable analogue output for feedback loop - Galvanic isolation between input and output - Serial communication via MODbus

- User definable measuring range Range extent: 10uS - 100mS Cell constant: 0.01 - 10 - Temperature compensation via Pt100/Pt1000 sensor - Readout of conductivity, temperature - Analogue output 0(4) - 20mA for conductivity or oC - Invertable analogue output for feedback loop - Galvanic isolation between input and output - Serial communication via MODbus

- 4 analogue inputs, 0(4)-20mA or 0-10V - Input 1 is retransmitted via the analogue output - Analogue output 0(4) - 20mA - Serial communication via MODbus - Transfers up to 4 measurements pr. module - Galvanic isolation between input and output

- Controller with 2 independent limits - Analogue input: 0(4) - 20mA - Zoom function: 100% at 10 - 20mA - 2 relay outputs - Optional max. or min. limits - Start timer and respons timers - Hysteresis function on both limits - Digital inputs: Optional active high or low